Podrobnosti návrhu

Číslo:62D/2280/CDV - ISO 80601-2-56 ED3
Zdroj:ISO\IEC\CEN\CENELEC
Komise:IEC/SC 62D
Název komise:Particular medical equipment, software, and systems
Návrh uveřejněn:16.12.2025
K připomínkám do:13.02.2026
Oblast zaměření:Řízení procesů
Contact email:aslana(at)agentura-cas.cz
Anotace:

IEC 60601-1:2005+AMD1:2012, 1.1 is replaced by:

This document applies to the basic safety and essential performance of a clinical thermometer in combination with its accessories, hereafter referred to as ME equipment. This document specifies the general and technical requirements for electrical clinical thermometers. This document only applies to all electrical clinical thermometers that are used for measuring the temperature of patients.

This document is also applicable to those accessories intended by their manufacturer to be connected to the clinical thermometer, where the characteristics of those accessories can affect the basic safety or essential performance of the clinical thermometer.

EXAMPLE 1         Probes, probe cable extenders and probe covers.

Clinical thermometers can be equipped with interfaces to accommodate secondary indicators, printing equipment, and other auxiliary equipment to create ME systems. This document does not apply to auxiliary equipment.

ME equipment that measures the temperature of a patient is within the scope of this document.

EXAMPLE 2        ME equipment using accessories such as a pulmonary artery catheter for the determination of cardiac output by thermodilution is in the scope of this document if it displays a temperature.

EXAMPLE 3        ME equipment using accessories such as a Foley catheter that includes a temperature probe is in the scope of this document.

This document does not specify requirements for screening thermographs intended to be used for the individual non-invasive human febrile temperature screening of humans under indoor environmental conditions, whose laboratory accuracy requirements are described in IEC 80601‑2‑59[4].

If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.

Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012, 7.2.13 and 8.4.1.

NOTE              Additional information can be found in IEC 60601–1:2005+AMD1:2012, 4.2.